FDAFDA · Alimentos y medicamentos·N.º oficial Z-2550-2026
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Publicado el 8 de julio de 2026 · Inicio del retiro: 21 de mayo de 2026
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Baxter Healthcare Corporation
- Distribuidor
- Worldwide distribution - US Nationwide and the country of Canada.
- Lote
- UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States