Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2551-2026

Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device

Publicado el 8 de julio de 2026 · Inicio del retiro: 21 de mayo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Baxter Healthcare Corporation
Distribuidor
Worldwide distribution - US Nationwide and the country of Canada.
Lote
UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.