Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2654-2025

Biograph mMR. Model Number: 10433372.

Publicado el 8 de octubre de 2025 · Inicio del retiro: 28 de agosto de 2025

Resumen

FDA registra el retiro Z-2654-2025 de dispositivos médicos de Clase I, publicado el 8 de octubre de 2025 e iniciado el 28 de agosto de 2025. Empresa responsable: Siemens Medical Solutions USA, Inc; Estado: En curso; Cantidad reportada: 27 units; Lote: Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Siemens Medical Solutions USA, Inc
Distribuidor
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Lote
Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Biograph mMR. Model Number: 10433372. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2654-2025 el 8 de octubre de 2025. Peligro reportado: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter