The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Publicado el 24 de septiembre de 2014 · Inicio del retiro: 1 de septiembre de 2011
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software. When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images. Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software. Philips will issue a software up
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Philips Medical Systems (Cleveland) Inc
- Distribuidor
- Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden.
- Lote
- 91003, 30018, 95504, 300004, 300030, 300010, 95130, 95157
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States