Alaris System PC Unit Model 8015, a modular infusion pump and monitoring system

Publicado el 12 de agosto de 2020 · Inicio del retiro: 30 de junio de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
CareFusion 303, Inc.
Distribuidor
Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
Lote
During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers. US and OUS serial numbers:
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States