Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2755-2017

Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027

Publicado el 2 de agosto de 2017 · Inicio del retiro: 7 de junio de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Navigation, Inc.
Distribuidor
US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX
Lote
Lot Numbers: 211158059, 211560973, 211959879, 12473141, 212707035, 211158151, 211594328, 211977586, 212503634, 212745660, 211158152, 211599898, 211987873, 212504463, 212796559, 211181212, 211628257, 212127654, 212549456, 212912501, 211181547,211908695, 212141557, 212553428, 211184941, 211916062, 212148368, 212632884,211555011, 211926889, 212434500, 212647830, 211555017, 211950736, 212438405, 212701679
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States