Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
Publicado el 5 de agosto de 2026 · Inicio del retiro: 8 de mayo de 2026
Resumen
FDA registra el retiro Z-2822-2026 de dispositivos médicos de Clase II, publicado el 5 de agosto de 2026 e iniciado el 8 de mayo de 2026. Empresa responsable: Instrumentation Laboratory; Estado: En curso; Cantidad reportada: 4 US; Lote: GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Instrumentation Laboratory
- Distribuidor
- US Nationwide distribution.
- Lote
- GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)21 de octubre de 2026
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic21 de octubre de 2026
- FDAGEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 0007741501012 de agosto de 2026
- FDAGEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 0005543001112 de agosto de 2026
- FDAGEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog5 de agosto de 2026
- FDABrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood5 de agosto de 2026
Preguntas sobre este retiro
¿Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer fue retirado del mercado?
Sí. FDA publicó el retiro Z-2822-2026 el 5 de agosto de 2026. Peligro reportado: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter