Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2855-2026

Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11

Publicado el 12 de agosto de 2026 · Inicio del retiro: 23 de marzo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
THERANICA BIOELECTRONICS LTD
Distribuidor
The product's entry point to the US is a 3PL warehouse from which it is being shipped to DME distributors or specialty pharmacy who perform the shipment to the end-user
Lote
Model/Catalog Number: FGD000100. UDI: 01072900175781181125120221NI2544-0300673 Software Version: FW Version 4.1.2.0, App version 7.28.0.11 GTIN: 7290017578118. Serial Number: NI2544-0300673.
Clasificación de riesgo
Class II
Área (FDA)
device
País
Israel

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta