munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
Publicado el 12 de agosto de 2026 · Inicio del retiro: 19 de junio de 2026
Resumen
FDA registra el retiro Z-2856-2026 de dispositivos médicos de Clase II, publicado el 12 de agosto de 2026 e iniciado el 19 de junio de 2026. Empresa responsable: MUNEVO GMBH; Estado: En curso; Cantidad reportada: 19; Lote: Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- MUNEVO GMBH
- Distribuidor
- Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
- Lote
- Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Germany
Retiros relacionados
- FDANerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.1112 de agosto de 2026
- FDABrand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.112 de agosto de 2026
- FDABrand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1,12 de agosto de 2026
- FDABrand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number:12 de agosto de 2026
- FDABrand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version:12 de agosto de 2026
- FDAAnti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System12 de agosto de 2026
Preguntas sobre este retiro
¿munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System fue retirado del mercado?
Sí. FDA publicó el retiro Z-2856-2026 el 12 de agosto de 2026. Peligro reportado: A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter