Philips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondary display control unit for the Expression MRI Patient Monitoring System and the Expression MR200 MRI Patient Monitoring System.
Publicado el 28 de septiembre de 2016 · Inicio del retiro: 14 de septiembre de 2014
Resumen
FDA registra el retiro Z-2867-2016 de dispositivos médicos de Clase II, publicado el 28 de septiembre de 2016 e iniciado el 14 de septiembre de 2014. Empresa responsable: Invivo Corporation; Estado: Finalizado; Cantidad reportada: 639 units; Lote: Serial numbers US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200410, US23200413, US23200415, US23200416, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434, US34600510, US34600511, US34600512, US34600513, US34600514, US34600515, US34600516, US34600517, US34600518, US34600519, US34600520, US34600521, US34600522, US34600523, US34600524, US34600525, US34600526, US34600528, US34600529, US34600530, US34600531, US34600532, US34600533, US34600534, US34600535, US34600536, US34600537, US34600538, US34600539, US34600540, US34600541, US34600542, US34600543, US34600544, US34600545, US34600546, US34600547, US34600548, US34600549, US34600550, US34600551, US34600552, U. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Frozen Display Numerics and Disabled Menu Keys after extended run time. This customer notification was sent August 22, 2014.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Invivo Corporation
- Distribuidor
- Worldwide distribution. US nationwide, Australia, Austria, Denmark, France, Germany, Iceland, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and England
- Lote
- Serial numbers US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200410, US23200413, US23200415, US23200416, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434, US34600510, US34600511, US34600512, US34600513, US34600514, US34600515, US34600516, US34600517, US34600518, US34600519, US34600520, US34600521, US34600522, US34600523, US34600524, US34600525, US34600526, US34600528, US34600529, US34600530, US34600531, US34600532, US34600533, US34600534, US34600535, US34600536, US34600537, US34600538, US34600539, US34600540, US34600541, US34600542, US34600543, US34600544, US34600545, US34600546, US34600547, US34600548, US34600549, US34600550, US34600551, US34600552, U
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMR Coils The MR Coil is intended to be used in conjunction with a Magnetic Resonance Scanner to produce28 de octubre de 2020
- FDAMonitor for the Functional Imaging System for Magnetic Resonance Imaging System11 de marzo de 2020
- FDATECNIS 1-Piece IOL with the TECNIS iTec Preloaded Delivery System, Model No. PCB00 The product is indicated28 de septiembre de 2016
- FDAPHS Spinal Tray 24G Gertie Marx, Catalogue No. 899379. Intended to provide licensed medical professionals with28 de septiembre de 2016
- FDAPhilips Expression MR200 MRI Patient Monitoring System22 de octubre de 2014
- FDAPhilips Expression MR200 MRI Patient Monitoring System Model 8661208 de octubre de 2014
Preguntas sobre este retiro
¿Philips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondary display control unit for the Expression MRI Patient Monitoring System and the Expression MR200 MRI Patient Monitoring System. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2867-2016 el 28 de septiembre de 2016. Peligro reportado: Frozen Display Numerics and Disabled Menu Keys after extended run time. This customer notification was sent August 22, 2014.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter