PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.
Publicado el 30 de septiembre de 2020 · Inicio del retiro: 31 de julio de 2020
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR¿ Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- MicroPort Orthopedics Inc.
- Distribuidor
- worldwide, except China and Japan
- Lote
- All lots/serial numbers
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States