Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2956-2024

Impella CP Pump Set, Japan; Product Code: 0048-0034-JP;

Publicado el 11 de septiembre de 2024 · Inicio del retiro: 5 de agosto de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Abiomed, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
Lote
Product Code: 0048-0034-JP; UDI-DI: 00813502011463; Serial Numbers: 324303 323153 328083 328085; Batch Numbers: 2022079852 2022079846 2022083745 2022083747
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States