Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Publicado el 2 de septiembre de 2026 · Inicio del retiro: 4 de agosto de 2026
Resumen
FDA registra el retiro Z-3014-2026 de dispositivos médicos de Clase II, publicado el 2 de septiembre de 2026 e iniciado el 4 de agosto de 2026. Empresa responsable: Canon Medical System, USA, INC.; Estado: En curso; Cantidad reportada: 10 systems; Lote: Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Canon Medical System, USA, INC.
- Distribuidor
- U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
- Lote
- Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000F4 de noviembre de 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C4 de noviembre de 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V4 de noviembre de 2026
- FDAAlphenix INFX-8000C/B514 de octubre de 2026
- FDAMidmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin2 de septiembre de 2026
- FDAStructur 2 SC Temporary Resin Cartridge Refill 75Gm, Product Codes: 1) 114-3933, Mfr Product Code 1479; Lots2 de septiembre de 2026
Preguntas sobre este retiro
¿Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. fue retirado del mercado?
Sí. FDA publicó el retiro Z-3014-2026 el 2 de septiembre de 2026. Peligro reportado: Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter