Optima MR450 1.5T Product Usage: 1 .5T Brivo MR355 and 1.5T Optima MR360 are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1 .5T Brivo MR355 and I .5T Optima MR360 reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Publicado el 23 de agosto de 2017 · Inicio del retiro: 11 de mayo de 2017
Resumen
FDA registra el retiro Z-3022-2017 de dispositivos médicos de Clase II, publicado el 23 de agosto de 2017 e iniciado el 11 de mayo de 2017. Empresa responsable: GE Medical Systems, LLC; Estado: Finalizado; Cantidad reportada: 68 units in total; Lote: Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Medical Systems, LLC
- Distribuidor
- US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
- Lote
- Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASenographe Pristina 3D11 de noviembre de 2026
- FDASenographe Pristina11 de noviembre de 2026
- FDASenographe Pristina11 de noviembre de 2026
- FDAOptima XR240amx, Mobile Digital Radiographic System22 de julio de 2026
- FDAVNS Therapy® AspireHC® Generator, VNS Therapy® AspireSR® Generator. Model No. 106. Product Usage: United23 de agosto de 2017
- FDAOptima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner23 de agosto de 2017
Preguntas sobre este retiro
¿Optima MR450 1.5T Product Usage: 1 .5T Brivo MR355 and 1.5T Optima MR360 are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the 1 .5T Brivo MR355 and I .5T Optima MR360 reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. fue retirado del mercado?
Sí. FDA publicó el retiro Z-3022-2017 el 23 de agosto de 2017. Peligro reportado: Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter