FDAFDA · Alimentos y medicamentos·N.º oficial Z-3023-2017
Optima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Optima MR450 1.5T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Publicado el 23 de agosto de 2017 · Inicio del retiro: 11 de mayo de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- GE Medical Systems, LLC
- Distribuidor
- US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
- Lote
- Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378
- Clasificación de riesgo
- Class II