FDAFDA · Alimentos y medicamentos·N.º oficial Z-3047-2017
VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Publicado el 30 de agosto de 2017 · Inicio del retiro: 1 de junio de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Solta Medical Inc
- Distribuidor
- US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.
- Lote
- Reference No. 110-0021; Shipping P/N P007021-03; Device P/N P006982-03; Serial No. VAS10-0134, VAS13-0440, VAS12-0203, 000TWY.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?