Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-3053-2017

Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it

Publicado el 30 de agosto de 2017 · Inicio del retiro: 29 de junio de 2017

Resumen

FDA registra el retiro Z-3053-2017 de dispositivos médicos de Clase II, publicado el 30 de agosto de 2017 e iniciado el 29 de junio de 2017. Empresa responsable: Medtronic Navigation, Inc.; Estado: Finalizado; Cantidad reportada: 75 instruments; Lote: Lot numbers 17030728, 17022027, 17021426, 17013125, 17012724, 17011303, 16122767, 16122066, 16112965, and 16112164. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Medtronic Navigation, Inc.
Distribuidor
Worldwide Distribution - US (nationwide) and Internationally to Canada.
Lote
Lot numbers 17030728, 17022027, 17021426, 17013125, 17012724, 17011303, 16122767, 16122066, 16112965, and 16112164
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it fue retirado del mercado?

Sí. FDA publicó el retiro Z-3053-2017 el 30 de agosto de 2017. Peligro reportado: The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter