Medtronic Suction 9734308 Small AxiEM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Publicado el 30 de agosto de 2017 · Inicio del retiro: 29 de junio de 2017
Resumen
FDA registra el retiro Z-3054-2017 de dispositivos médicos de Clase II, publicado el 30 de agosto de 2017 e iniciado el 29 de junio de 2017. Empresa responsable: Medtronic Navigation, Inc.; Estado: Finalizado; Cantidad reportada: 60 instruments; Lote: Lot numbers 17022451,17011750, 16120349, 16082989, 16081888, 16071987, 16062713, and 16060712. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Medtronic Navigation, Inc.
- Distribuidor
- Worldwide Distribution - US (nationwide) and Internationally to Canada.
- Lote
- Lot numbers 17022451,17011750, 16120349, 16082989, 16081888, 16071987, 16062713, and 16060712
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMedtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly28 de octubre de 2026
- FDAMedtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly28 de octubre de 2026
- FDAMedtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation,30 de agosto de 2017
- FDAMedtronic Instrument Set EM ENT, Ref. #9733452 and #9733452-G02, reusable, RX. This kit contains 9733449,30 de agosto de 2017
Preguntas sobre este retiro
¿Medtronic Suction 9734308 Small AxiEM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it fue retirado del mercado?
Sí. FDA publicó el retiro Z-3054-2017 el 30 de agosto de 2017. Peligro reportado: The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter