Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators

Publicado el 20 de septiembre de 2017 · Inicio del retiro: 8 de agosto de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Leonhard Lang Medizintechnik GmbH
Distribuidor
Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
Lote
Model Number DF59N (lot numbers 41219-0778, 50403-0772, 41029-0770), DF59NC (lot numbers 50211-0974, 50402-0974, 40806-0972, 41203-0973, 50116-0975, 50527-0974). Expiration date 09/2017 to 05/31/2018.
Clasificación de riesgo
Class I
Área (FDA)
device
País
Austria