BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X
Publicado el 2 de octubre de 2024 · Inicio del retiro: 2 de julio de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Argon Medical Devices, Inc
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.
- Lote
- REF/Box UDI-DI/Unit UDI-DI/Lot(Expiration Date): 360-1080-01/20886333004373/00886333004379/11563436(22-Mar-2027), 11568010(16-Apr-2027), 11568790(1-May-2027), 11569552(7-May-2027), 11572560(20-May-2027) 360-1080-02/20886333224009/00886333224005/11563678(22-Mar-27), 11562475(05-Mar-27), 11562476(14-Mar-27), 11566932(08-Apr-27), 11567139(11-Apr-27), 11568361(16-Apr-27), 11569013(23-Apr-27), 11569844(29-Apr-27), 11571094(02-May-27), 11572044(08-May-27), 11572886(26-May-27); 360-1080-03/20886333224016/00886333224012/11569845(22-Apr-27), 11562477(04-Mar-27), 11564861(26-Mar-27), 11567888(12-Apr-27), 11568554(18-Apr-27), 11570075(04-May-27); 360-1580-01/20886333004359/00886333004355/11563438(21-Mar-2027), 11563439(19-Mar-2027); 360-1580-02/20886333224023/00886333224029/11562478(14-Mar-27), 11563440(18-Mar-27), 11563682(20-Mar-27), 11565238(03-Apr-27), 11566882(09-Apr-27), 11570132(01-May-27), 11570814(09-May-27), 11573339(20-May-27); 360-1580-03/20886333224030/00886333224036/11570288(02-May-27), 11563442(20-Mar-27), 11566173(03-Apr-27), 11566883(10-Apr-27), 11568153(16-Apr-27), 11570813(08-May-27); 360-2080-01/20886333004342/00886333004348/11563564(26-Mar-2027), 11566036(8-Apr-2027), 11567992(16-Apr-2027), 11570142(1-May-27), 11570863(8-May-27); 360-2080-02/20886333224047/00886333224043/11562964(07-Mar-27), 11563922(15-Mar-27), 11564189(22-Mar-27); 360-2080-03/20886333224054/00886333224050/11563443(15-Mar-27), 11566936(11-Apr-27), 11567889(22-Apr-2027); 370-1080-01/20886333004380/00886333004386/11564795(26-Mar-27); 370-1080-02/20886333224061/00886333224067/11562608(05-Mar-27), 11562748(04-Mar-27), 11564843(26-Mar-27), 11566037(08-Apr-27), 11571101(30-Apr-27), 11571360(06-May-27); 370-1080-03/20886333224078/00886333224074/11564844(22-Mar-27), 11566690(16-Apr-27), 11571100(01-May-27), 11573030(13-May-27); 370-1580-01/20886333004366/00886333004382/11564366(26-Mar-27); 370-1580-02/20886333224085/00886333224081/11571824(06-May-27), 11562114(07-Mar-27), 11566038(08-Apr-27), 11571824(06-May-27); 370-1580-03/20886333224092/00886333224098/11569414(22-Apr-27), 11571104(01-May-27); 763114100x/20886333006643/00886333006649/11563673(11-May-2029), 11570356(23-Apr-2029), 11571058(4-Apr-2029), 11572456(29-May-2029), 11572868(13-May-2029), 11573300(22-May-2029); 763116100X/20886333006612/00886333006618/11562815(7-Mar-2029), 11571061(29-Apr-2029); 763116160X/20886333006605/00886333006601/11571063(1-May-2029); 763118100X/20886333006582/00886333006588/11564330(26-Mar-2029); 763118200X/20886333006650/00886333006656/11562085(5-Mar-2029), 11571065(4-APR-2029); 763120100X/20886333006568/00886333006564/11564331(26-Mar-2029); 763120160X/20886333006575/00886333006571/11565618(10-Apr-2029); 763418200X/20886333006803/00886333006809/11566213(10-Apr-2029); 763418250X/20886333006810/00886333006816/11564594(22-Mar-2029), 11565619(4-Apr-2029), 11572175(22-May-2029); 763114200X/20886333006629/00886333006625/11566106(4-Apr-2029)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States