FDAFDA · Alimentos y medicamentos·N.º oficial Z-3233-2017
Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).
Publicado el 4 de octubre de 2017 · Inicio del retiro: 24 de julio de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Sorin Group Italia SRL - CRF
- Distribuidor
- US Nationwide Distribution
- Lote
- All Serial Numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Italy
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.