Siemens RAPIDPoint® 500 system blood gas analyzer Product Usage: RAPIDPoint 500 System: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili
Published November 6, 2013 · Recall initiated: September 17, 2012
Summary
FDA records medical device Class II recall Z-0104-2014, published November 6, 2013 and initiated September 17, 2012. Recalling firm: Siemens Healthcare Diagnostics Inc; Status: Terminated; Reported quantity: 2152 units; Lot: All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
RAPIDPoint® 400 Series, RAPIDPoint® 500, or RAPIDLab® 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam,
- Lot
- All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;May 6, 2026
- FDARAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.August 6, 2025
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515November 13, 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPointJuly 3, 2024
- FDASiemens RAPIDPoint®¿ 405 system blood gas analyzer Product PUsage: These systems are intended for near-patientNovember 6, 2013
- FDASiemens RAPIDLab® 1240 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200November 6, 2013
Questions about this recall
Was Siemens RAPIDPoint® 500 system blood gas analyzer Product Usage: RAPIDPoint 500 System: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili recalled?
Yes. FDA published recall Z-0104-2014 on November 6, 2013. Reported hazard: RAPIDPoint® 400 Series, RAPIDPoint® 500, or RAPIDLab® 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
What should I do if I have this product?
Voluntary: Firm initiated · Letter