Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry Catalog Number:1360 Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Published January 16, 2013 · Recall initiated: November 21, 2012
Summary
FDA records medical device Class III recall Z-0635-2013, published January 16, 2013 and initiated November 21, 2012. Recalling firm: Siemens Healthcare Diagnostics Inc; Status: Terminated; Reported quantity: 7865; Lot: Carton/Bottle Lot Number: C9093061B P9093061B C9094081B P9094081B C0097091B P0097091B C0100101C P0100101C C0103042F P0103042F C0106062B P0106062B C91 01022D P91 01022D CP9105052B P9105052B C9108092A P9108092A. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25®C (77®F).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
- Lot
- Carton/Bottle Lot Number: C9093061B P9093061B C9094081B P9094081B C0097091B P0097091B C0100101C P0100101C C0103042F P0103042F C0106062B P0106062B C91 01022D P91 01022D CP9105052B P9105052B C9108092A P9108092A
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;May 6, 2026
- FDARAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.August 6, 2025
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515November 13, 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPointJuly 3, 2024
- FDAProtandim is packaged in a 90cc cylindrical blue bottle with a silver cap and contains thirty caplets perJanuary 16, 2013
- FDASiemens CLINITEK Atlas Positive and Negative Control Strips for Urine Chemistry Catalog Number: 5019 When usedJanuary 16, 2013
Questions about this recall
Was Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry Catalog Number:1360 Chek-Stix® Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX® positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes recalled?
Yes. FDA published recall Z-0635-2013 on January 16, 2013. Reported hazard: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25®C (77®F).
What should I do if I have this product?
Voluntary: Firm initiated · Letter