FDAFDA · Food and drugs·Recall number Z-0638-2013
Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry (Distributed OUS Only) Catalog Number: 1360L Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Published January 16, 2013 · Recall initiated: November 21, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
- Lot
- Carton/Bottle Lot Number: C9094081C P9094081C C0100101A P0100101A C9101022C P9101022C
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Sell products like this?