ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.
Published February 10, 2016 · Recall initiated: January 11, 2016
Summary
FDA records medical device Class II recall Z-0740-2016, published February 10, 2016 and initiated January 11, 2016. Recalling firm: Horiba Instruments Inc; Status: Terminated; Reported quantity: 216; Lot: All Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Horiba Instruments Inc
- Distributor
- U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
- Lot
- All Lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determinationFebruary 10, 2016
- FDAABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the quantitative determination ofFebruary 10, 2016
- FDAHemosIL Plasminogen, Part No. 0020009000, IVD. Instrumentation Laboratory Company. Kit contents: 2 x 2.5 mLFebruary 10, 2016
- FDAPS-500 Scope Warmer; 10 units per box. Pre-heating scope for laparoscopy surgery.February 10, 2016
- FDABaxter, RAPIDFILL Syringe Strip, 10 ml. For use with RAPIDFILL Automated Syringe Filler (ASF)February 10, 2016
- FDALite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D TheDecember 31, 2014
Questions about this recall
Was ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry. recalled?
Yes. FDA published recall Z-0740-2016 on February 10, 2016. Reported hazard: N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
What should I do if I have this product?
Voluntary: Firm initiated · Letter