Back to search
FDAFDA · Food and drugs·Recall number Z-0742-2016

ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.

Published February 10, 2016 · Recall initiated: January 11, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Horiba Instruments Inc
Distributor
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Lot
All Lots
Risk classification
Class II
Area (FDA)
device
Country
United States