BriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, Qty 10, (01)15055788723001 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.
Published January 22, 2020 · Recall initiated: November 27, 2019
Summary
FDA records medical device Class II recall Z-0801-2020, published January 22, 2020 and initiated November 27, 2019. Recalling firm: Flexicare Medical Ltd.; Status: Terminated; Reported quantity: 1,024,890 devices; Lot: Part Number/Lot Number 040-713U 180700030 I CCW 040-713U 180900099 I CCW 040-713U 180901353 I CCW 040-713U 181000110 I CCW 040-713U 181000113 I CCW 040-713U 181100001 I CCW 040-713U 181100002 I CCW 040-713U 181200005 I CCW. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Flexicare Medical Ltd.
- Distributor
- US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
- Lot
- Part Number/Lot Number 040-713U 180700030 I CCW 040-713U 180900099 I CCW 040-713U 180901353 I CCW 040-713U 181000110 I CCW 040-713U 181000113 I CCW 040-713U 181100001 I CCW 040-713U 181100002 I CCW 040-713U 181200005 I CCW
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABriteBlade Pro Single-Use Fiber Optic Handle and Blade Mac 3, CE 0120, Do not Re-use, Rx Only, REF, 040-333U,January 22, 2020
- FDABriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, QtyJanuary 22, 2020
- FDABriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF:January 22, 2020
- FDAQuickDraw Venous Cannula, REF: QD22, UDI: 00690103182699; and REF: QD25, UDI: 00690103182705January 22, 2020
- FDAAirway Gas Option N-CAiOJanuary 22, 2020
- FDACARESCAPE Respiratory Module E-sCAiOVJanuary 22, 2020
Questions about this recall
Was BriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, Qty 10, (01)15055788723001 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis. recalled?
Yes. FDA published recall Z-0801-2020 on January 22, 2020. Reported hazard: The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit