BriteBlade Pro Single-Use Fiber Optic Handle and Blade Mac 3, CE 0120, Do not Re-use, Rx Only, REF, 040-333U, Sterile EO, Qty 10, (01)15055788724138(17)230801(10)180900072 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.
Published January 22, 2020 · Recall initiated: November 27, 2019
Summary
FDA records medical device Class II recall Z-0803-2020, published January 22, 2020 and initiated November 27, 2019. Recalling firm: Flexicare Medical Ltd.; Status: Terminated; Reported quantity: 1,024,890 devices; Lot: Part Number/Lot Number 040-333U 170900328 I OGH 040-333U 180800461 I OGH 040-333U 180900072 I OGH 040-333U 180900104 I OGH 040-333U 181000107 I OGH 040-333U 181100013 I OGH 040-333U 181100097 I OGH 040-333U 181200024 I OGH. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Flexicare Medical Ltd.
- Distributor
- US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
- Lot
- Part Number/Lot Number 040-333U 170900328 I OGH 040-333U 180800461 I OGH 040-333U 180900072 I OGH 040-333U 180900104 I OGH 040-333U 181000107 I OGH 040-333U 181100013 I OGH 040-333U 181100097 I OGH 040-333U 181200024 I OGH
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, QtyJanuary 22, 2020
- FDABriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF:January 22, 2020
- FDABriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, QtyJanuary 22, 2020
- FDASiemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: AngiographyJanuary 22, 2020
- FDAQuickDraw Venous Cannula, REF: QD22, UDI: 00690103182699; and REF: QD25, UDI: 00690103182705January 22, 2020
- FDAMicrobiologics QC Sets and Panels, Gram-Positive Blood Culture Control Panel (Inactivated Pellet), Catalog #January 22, 2020
Questions about this recall
Was BriteBlade Pro Single-Use Fiber Optic Handle and Blade Mac 3, CE 0120, Do not Re-use, Rx Only, REF, 040-333U, Sterile EO, Qty 10, (01)15055788724138(17)230801(10)180900072 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis. recalled?
Yes. FDA published recall Z-0803-2020 on January 22, 2020. Reported hazard: The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit