LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects.
Published February 27, 2013 · Recall initiated: January 4, 2013
Summary
FDA records medical device Class I recall Z-0830-2013, published February 27, 2013 and initiated January 4, 2013. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 162 total units (US) and 102 units total (OUS); Lot: Part number: 198720024 and Lot numbers: 113812, 123509, 140578, 156210, 171659, 171660, 171661, 171662, 171663, 123509A, B34GE1000, B34GEA000, B4EGR1000, B4EGRA000, B4EGRB000, B4XFL1000, B4XFLA000, B5SH11000, B5SH1A000, B5SH1B, C74J11000, D14KG1000, E4RF21000, EN8CY1000, EN8CYA000, FC1MB1, and FC1MBA. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South Africa and Canada including Veteran Administration hospitals.
- Lot
- Part number: 198720024 and Lot numbers: 113812, 123509, 140578, 156210, 171659, 171660, 171661, 171662, 171663, 123509A, B34GE1000, B34GEA000, B4EGR1000, B4EGRA000, B4EGRB000, B4XFL1000, B4XFLA000, B5SH11000, B5SH1A000, B5SH1B, C74J11000, D14KG1000, E4RF21000, EN8CY1000, EN8CYA000, FC1MB1, and FC1MBA.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212March 29, 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312March 29, 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDALPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is anFebruary 27, 2013
- FDALPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is anFebruary 27, 2013
Questions about this recall
Was LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects. recalled?
Yes. FDA published recall Z-0830-2013 on February 27, 2013. Reported hazard: DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.
What should I do if I have this product?
Voluntary: Firm initiated · Letter