Back to search
FDAFDA · Food and drugs·Recall number Z-0832-2020

PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700

Published January 29, 2020 · Recall initiated: December 2, 2019

Summary

FDA records medical device Class II recall Z-0832-2020, published January 29, 2020 and initiated December 2, 2019. Recalling firm: Polymer Technology Systems, Inc.; Status: Terminated; Reported quantity: 44,448 total; Lot: All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Polymer Technology Systems, Inc.
Distributor
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
Lot
All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700 recalled?

Yes. FDA published recall Z-0832-2020 on January 29, 2020. Reported hazard: One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

What should I do if I have this product?

Voluntary: Firm initiated · Letter