FDAFDA · Food and drugs·Recall number Z-0832-2020
PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
Published January 29, 2020 · Recall initiated: December 2, 2019
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Polymer Technology Systems, Inc.
- Distributor
- The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
- Lot
- All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States