PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
Published January 29, 2020 · Recall initiated: December 2, 2019
Summary
FDA records medical device Class II recall Z-0832-2020, published January 29, 2020 and initiated December 2, 2019. Recalling firm: Polymer Technology Systems, Inc.; Status: Terminated; Reported quantity: 44,448 total; Lot: All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Polymer Technology Systems, Inc.
- Distributor
- The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
- Lot
- All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308January 29, 2020
- FDAGendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)January 29, 2020
- FDAPTS Panels CHOL+HDL+GLU Test Strips, Reference Number 5412, for use on CardioChek PA or CardioChek PlusJanuary 1, 2020
- FDALipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry ScheinMay 29, 2019
- FDACHOL+HDL+GLU test strips, REF 2412May 29, 2019
- FDAPTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcareJune 5, 2013
Questions about this recall
Was PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700 recalled?
Yes. FDA published recall Z-0832-2020 on January 29, 2020. Reported hazard: One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
What should I do if I have this product?
Voluntary: Firm initiated · Letter