Back to search
FDAFDA · Food and drugs·Recall number Z-1372-2019

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Published May 29, 2019 · Recall initiated: January 16, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Polymer Technology Systems, Inc.
Distributor
Domestic distribution nationwide.
Lot
Q802 Q811 Q813 Q816 Q820 Q823 Q831 Q903
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card