Back to search
FDAFDA · Food and drugs·Recall number Z-1373-2019

CHOL+HDL+GLU test strips, REF 2412

Published May 29, 2019 · Recall initiated: January 16, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Polymer Technology Systems, Inc.
Distributor
Domestic distribution nationwide.
Lot
I808 I809 I811 I812 I814
Risk classification
Class II
Area (FDA)
device
Country
United States