PTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcare professionals and individuals with diabetes to measure glucose in whole blood
Published June 5, 2013 · Recall initiated: April 4, 2013
Summary
FDA records medical device Class II recall Z-1403-2013, published June 5, 2013 and initiated April 4, 2013. Recalling firm: Polymer Technology Systems, Inc.; Status: Terminated; Reported quantity: 7048 total, 3499 in USA; Lot: U1117. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Internal investigation demonstrated under-recovery for glucose when tested against a reference method. This could result in a low to moderate health risk.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Polymer Technology Systems, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide including Puerto Rico and countries of: UK, CHL, ENG, BEL, ITA, TKY, BEL, LIT, LTV, EGY, SPG, HOK, SWE, MEX, UAE, COR, ISR, MAL, AUS, POR, POL, SLN and STQ.
- Lot
- U1117
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, ReferenceJanuary 29, 2020
- FDAPTS Panels CHOL+HDL+GLU Test Strips, Reference Number 5412, for use on CardioChek PA or CardioChek PlusJanuary 1, 2020
- FDALipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry ScheinMay 29, 2019
- FDACHOL+HDL+GLU test strips, REF 2412May 29, 2019
- FDAIntegrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibilityJune 5, 2013
- FDACanon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to beJune 5, 2013
Questions about this recall
Was PTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcare professionals and individuals with diabetes to measure glucose in whole blood recalled?
Yes. FDA published recall Z-1403-2013 on June 5, 2013. Reported hazard: Internal investigation demonstrated under-recovery for glucose when tested against a reference method. This could result in a low to moderate health risk.
What should I do if I have this product?
Voluntary: Firm initiated · Letter