AF w/AS 550 x 1-14ML/HR 2ML-15LO CPNB, Product code MVBT15XL-CPNB, Arrow AutoFuser Disposable Pain Control Pump
Published March 6, 2019 · Recall initiated: October 2, 2018
Summary
FDA records medical device Class II recall Z-0954-2019, published March 6, 2019 and initiated October 2, 2018. Recalling firm: Arrow International Inc; Status: Terminated; Reported quantity: 19; Lot: 74B1702339 74H1601294 74D1701102 74H1700162 74D1702253 74H1701109 74E1701025 74H1701794 74F1700780 74J1700192 74F1701946 74J1700678 74G1600394 74J1701748 74G1700546 74K1702039 74G1701177 74L1700931 74G1702054. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Arrow International Inc
- Distributor
- Nationwide distribution.
- Lot
- 74B1702339 74H1601294 74D1701102 74H1700162 74D1702253 74H1701109 74E1701025 74H1701794 74F1700780 74J1700192 74F1701946 74J1700678 74G1600394 74J1701748 74G1700546 74K1702039 74G1701177 74L1700931 74G1702054
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500JJuly 8, 2020
- FDAArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701July 8, 2020
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500July 8, 2020
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400July 8, 2020
- FDAAF w/AS 550 x 1-14 ml/hr, 5ml30 LOCPNB, Product code MVBX30XLCPNB, Arrow AutoFuser Disposable Pain ControlMarch 6, 2019
- FDAAF w/AS 550 x 1-14ML/HR 2ML-30LO CPNB, Product code MVBT30XL CPNB, Arrow AutoFuser Disposable Pain ControlMarch 6, 2019
Questions about this recall
Was AF w/AS 550 x 1-14ML/HR 2ML-15LO CPNB, Product code MVBT15XL-CPNB, Arrow AutoFuser Disposable Pain Control Pump recalled?
Yes. FDA published recall Z-0954-2019 on March 6, 2019. Reported hazard: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
What should I do if I have this product?
Voluntary: Firm initiated · Letter