AF w/AS 550 x 1-14ML/HR 5ML-60LO CPNB, Product code MVBX60XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Published March 6, 2019 · Recall initiated: October 2, 2018
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Arrow International Inc
- Distributor
- Nationwide distribution.
- Lot
- 74E1702443
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States