REF 623401 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time.
Published March 19, 2014 · Recall initiated: February 24, 2014
Summary
FDA records medical device Class II recall Z-1203-2014, published March 19, 2014 and initiated February 24, 2014. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 129; Lot: 249649, 102928, 102929, 109312, 128472, 142839, 142840, 142841, 162485, 196050, 196051, 196052, 211382, 211387, 213581, 227387, 231687, 249644, 249647, 269701, 269702, 288346, 298769, 298770, 298772, 306503, 306505, 329328, 338142, 346434, 354575, 360633, 360634, 366429, 366430, 373254, 378641, 378642, 378644, 402462, 403004, 403005, 403007, 403008, 421300, 438246, 438247, 438248, 438249, 452447, 456738, 456740, 466369, 466370, 466371, 058401LA, 137401LA, D2WBG4, D37FW4, D37GP4, DV6LY4, DY5EC4, DY5ED4, E1VHM4, E3SDW4, E6CE94, E82E14, EA8EH4, EF5BS4, EJ6G34, EW6HT4, FJ3D14, FJ3D24, and FJ3DY4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM® Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Australia, Austria, China, Czech Republic, Denmark, Dubai / UAE, Estonia, France, Germany, Hungary, Iberia, India, Ireland, Israel, Italy, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russia, Singapore, Slovenia, Spain, South Africa, Sweden, Switzerland, Turkey, United Kingdom, and Vietnam.
- Lot
- 249649, 102928, 102929, 109312, 128472, 142839, 142840, 142841, 162485, 196050, 196051, 196052, 211382, 211387, 213581, 227387, 231687, 249644, 249647, 269701, 269702, 288346, 298769, 298770, 298772, 306503, 306505, 329328, 338142, 346434, 354575, 360633, 360634, 366429, 366430, 373254, 378641, 378642, 378644, 402462, 403004, 403005, 403007, 403008, 421300, 438246, 438247, 438248, 438249, 452447, 456738, 456740, 466369, 466370, 466371, 058401LA, 137401LA, D2WBG4, D37FW4, D37GP4, DV6LY4, DY5EC4, DY5ED4, E1VHM4, E3SDW4, E6CE94, E82E14, EA8EH4, EF5BS4, EJ6G34, EW6HT4, FJ3D14, FJ3D24, and FJ3DY4
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212March 29, 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312March 29, 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDAMar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.March 19, 2014
- FDAREF 623401 R, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Medium, SterileMarch 19, 2014
Questions about this recall
Was REF 623401 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time. recalled?
Yes. FDA published recall Z-1203-2014 on March 19, 2014. Reported hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM® Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail