Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), Evita V500 and /or Babylog VN500 Ventilators, used for ventilation during short term transport of patients within a hospital.
Published April 9, 2014 · Recall initiated: February 7, 2014
Summary
FDA records medical device Class I recall Z-1298-2014, published April 9, 2014 and initiated February 7, 2014. Recalling firm: Draeger Medical Systems, Inc.; Status: Terminated; Reported quantity: 1544; Lot: all lots. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Draeger Medical Systems, Inc.
- Distributor
- Nationwide Distribution
- Lot
- all lots
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21,July 3, 2024
- FDADraeger Jaundice Meter JM-103 The device is intended for use in hospitals, clinics or doctor s offices under aJune 13, 2018
- FDADraeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833April 16, 2014
- FDAAsian Taste (Brand) Green Tea Pumpkin Seeds, 8 oz., UPC 673367418186.April 9, 2014
- FDARayStation Radiation Therapy Treatment Planning System, models 2.5, 3.0, 3.5 and 4.0; a software systemApril 9, 2014
- FDADraeger Infinity Acute Care System Monitoring SolutionJanuary 22, 2014
Questions about this recall
Was Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), Evita V500 and /or Babylog VN500 Ventilators, used for ventilation during short term transport of patients within a hospital. recalled?
Yes. FDA published recall Z-1298-2014 on April 9, 2014. Reported hazard: The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter