Draeger Infinity Acute Care System Monitoring Solution
Published January 22, 2014 · Recall initiated: December 10, 2013
Summary
FDA records medical device Class II recall Z-0714-2014, published January 22, 2014 and initiated December 10, 2013. Recalling firm: Draeger Medical Systems, Inc.; Status: Terminated; Reported quantity: 439; Lot: with software versions VG2.0.3 and higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Draeger Medical Systems, Inc.
- Distributor
- US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
- Lot
- with software versions VG2.0.3 and higher.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21,July 3, 2024
- FDADraeger Jaundice Meter JM-103 The device is intended for use in hospitals, clinics or doctor s offices under aJune 13, 2018
- FDADraeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833April 16, 2014
- FDADraeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), EvitaApril 9, 2014
- FDApO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units,January 22, 2014
- FDAAnspach MCA Cutting Burr, DFU Rx Only Designed for cutting and shaping delicate bone, primarily in otologyJanuary 22, 2014
Questions about this recall
Was Draeger Infinity Acute Care System Monitoring Solution recalled?
Yes. FDA published recall Z-0714-2014 on January 22, 2014. Reported hazard: After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
What should I do if I have this product?
Voluntary: Firm initiated · Letter