Draeger Infinity Acute Care System Monitoring Solution

Published January 22, 2014 · Recall initiated: December 10, 2013

Summary

FDA records medical device Class II recall Z-0714-2014, published January 22, 2014 and initiated December 10, 2013. Recalling firm: Draeger Medical Systems, Inc.; Status: Terminated; Reported quantity: 439; Lot: with software versions VG2.0.3 and higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Draeger Medical Systems, Inc.
Distributor
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Lot
with software versions VG2.0.3 and higher.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Draeger Infinity Acute Care System Monitoring Solution recalled?

Yes. FDA published recall Z-0714-2014 on January 22, 2014. Reported hazard: After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

What should I do if I have this product?

Voluntary: Firm initiated · Letter