Back to search
FDAFDA · Food and drugs·Recall number Z-0714-2014

Draeger Infinity Acute Care System Monitoring Solution

Published January 22, 2014 · Recall initiated: December 10, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Draeger Medical Systems, Inc.
Distributor
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Lot
with software versions VG2.0.3 and higher.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card