Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Published April 16, 2014 · Recall initiated: March 5, 2014
Summary
FDA records medical device Class I recall Z-1440-2014, published April 16, 2014 and initiated March 5, 2014. Recalling firm: Draeger Medical Systems, Inc.; Status: Terminated; Reported quantity: 3; Lot: Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Draeger Medical Systems, Inc.
- Distributor
- USA Nationwide Distribution in the states of WI and NY.
- Lot
- Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21,July 3, 2024
- FDADraeger Jaundice Meter JM-103 The device is intended for use in hospitals, clinics or doctor s offices under aJune 13, 2018
- FDAPhilips HeartStart XL+ Defibrillator/Monitor Model number: 861290, automatic external defibrillator.April 16, 2014
- FDAFAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended toApril 16, 2014
- FDADraeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), EvitaApril 9, 2014
- FDADraeger Infinity Acute Care System Monitoring SolutionJanuary 22, 2014
Questions about this recall
Was Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833 recalled?
Yes. FDA published recall Z-1440-2014 on April 16, 2014. Reported hazard: The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
What should I do if I have this product?
Voluntary: Firm initiated · Letter