Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Published April 16, 2014 · Recall initiated: March 5, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Draeger Medical Systems, Inc.
- Distributor
- USA Nationwide Distribution in the states of WI and NY.
- Lot
- Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States