Back to search
FDAFDA · Food and drugs·Recall number Z-1440-2014

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Published April 16, 2014 · Recall initiated: March 5, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Draeger Medical Systems, Inc.
Distributor
USA Nationwide Distribution in the states of WI and NY.
Lot
Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
Risk classification
Class I
Area (FDA)
device
Country
United States