Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
Published July 3, 2024 · Recall initiated: May 13, 2024
Summary
FDA records medical device Class II recall Z-2189-2024, published July 3, 2024 and initiated May 13, 2024. Recalling firm: Draeger Medical Systems, Inc.; Status: Ongoing; Reported quantity: 34,102 units; Lot: Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Draeger Medical Systems, Inc.
- Distributor
- US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T®rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
- Lot
- Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAZE for Vertix Solitaire - The AXIOM Vertix Solitaire is a component and has been specially designed forJuly 3, 2024
- FDACOBALT HF QUAD CRT-D MRI SureScan, Model Number DTPB2QQ, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDADraeger Jaundice Meter JM-103 The device is intended for use in hospitals, clinics or doctor s offices under aJune 13, 2018
- FDADraeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833April 16, 2014
- FDADraeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), EvitaApril 9, 2014
- FDADraeger Infinity Acute Care System Monitoring SolutionJanuary 22, 2014
Questions about this recall
Was Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23 recalled?
Yes. FDA published recall Z-2189-2024 on July 3, 2024. Reported hazard: Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter