AirLife Resuscitation, Infant, with mask, 40inch oxygen reservoir tubing, pressure-relief valve, manometer, PEEP valve Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.

Published May 16, 2018 · Recall initiated: April 27, 2017

Summary

FDA records medical device Class I recall Z-1580-2018, published May 16, 2018 and initiated April 27, 2017. Recalling firm: Vyaire Medical; Status: Terminated; Reported quantity: 198 units; Lot: Part Number: 2K8038M. Lots: 0001045566, 0001074625, 0001080151, 0001083196, 0001086176. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Vyaire Medical
Distributor
Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.
Lot
Part Number: 2K8038M. Lots: 0001045566, 0001074625, 0001080151, 0001083196, 0001086176
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AirLife Resuscitation, Infant, with mask, 40inch oxygen reservoir tubing, pressure-relief valve, manometer, PEEP valve Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own. recalled?

Yes. FDA published recall Z-1580-2018 on May 16, 2018. Reported hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.

What should I do if I have this product?

Voluntary: Firm initiated · Letter