Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.
Published April 22, 2020 · Recall initiated: January 21, 2020
Summary
FDA records medical device Class II recall Z-1717-2020, published April 22, 2020 and initiated January 21, 2020. Recalling firm: Sysmex America, Inc.; Status: Terminated; Reported quantity: 40 units; Lot: Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sysmex America, Inc.
- Distributor
- USA: FL, VT, IL.
- Lot
- Serial Numbers: 11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stainedApril 22, 2020
- FDAVACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.April 22, 2020
- FDAEndocuff Vision: models ARV 110, 120, 130, 140 Intended Use: ENDOCUFF VISION is a device attached to theApril 22, 2020
- FDAThe UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. TheFebruary 19, 2014
- FDASysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
- FDASysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
Questions about this recall
Was Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers. recalled?
Yes. FDA published recall Z-1717-2020 on April 22, 2020. Reported hazard: When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
What should I do if I have this product?
Voluntary: Firm initiated · Letter