Back to search
FDAFDA · Food and drugs·Recall number Z-2264-2013

ENDOPLUS; Precision Laparoscopic Instruments; REF 346VA; Description 5mm Atraumatic Grasper S/A, Standard; Non-sterile; Manufactured by Endoplus; 750 Tower Rd., Suite A, Mundelein, IL 60060 USA. Intended for grasping, dissecting, retraction, clamping in conjunction with the laparoscopic during laparoscopic surgery.

Published October 2, 2013 · Recall initiated: July 15, 2013

Summary

FDA records medical device Class II recall Z-2264-2013, published October 2, 2013 and initiated July 15, 2013. Recalling firm: Endoplus, Inc.; Status: Terminated; Reported quantity: 81 units; Lot: Item: 346VA; Lot Numbers: D3, E3, F3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Multiple complaints reporting that the moveable jaw has broken during laparoscopic procedures.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Endoplus, Inc.
Distributor
Distributed in the states of FL, TX, and VA.
Lot
Item: 346VA; Lot Numbers: D3, E3, F3
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ENDOPLUS; Precision Laparoscopic Instruments; REF 346VA; Description 5mm Atraumatic Grasper S/A, Standard; Non-sterile; Manufactured by Endoplus; 750 Tower Rd., Suite A, Mundelein, IL 60060 USA. Intended for grasping, dissecting, retraction, clamping in conjunction with the laparoscopic during laparoscopic surgery. recalled?

Yes. FDA published recall Z-2264-2013 on October 2, 2013. Reported hazard: Multiple complaints reporting that the moveable jaw has broken during laparoscopic procedures.

What should I do if I have this product?

Voluntary: Firm initiated · Letter