Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Published August 21, 2019 · Recall initiated: June 28, 2019
Summary
FDA records medical device Class II recall Z-2265-2019, published August 21, 2019 and initiated June 28, 2019. Recalling firm: Degania Silicone, Ltd.; Status: Terminated; Reported quantity: 2160; Lot: Lot Codes: P1851056 P1867059. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Degania Silicone, Ltd.
- Distributor
- Nationwide Distribution
- Lot
- Lot Codes: P1851056 P1867059
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Israel
Related recalls
- FDAGo-EZ Screw 6.5x38, Item Number 19655 - Product Usage: Accessories to the BioPro Go-Ez screw system and theAugust 21, 2019
- FDASOMATOM Drive, Model Number 10431700 - Product Usage: Computed tomography systems intended to generate andAugust 21, 2019
- FDAFoley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This isSeptember 26, 2018
- FDATemperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: ThisSeptember 26, 2018
- FDATemperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, singleSeptember 26, 2018
- FDATemperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use,September 26, 2018
Questions about this recall
Was Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions. recalled?
Yes. FDA published recall Z-2265-2019 on August 21, 2019. Reported hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail