Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Published September 26, 2018 · Recall initiated: June 21, 2018

Summary

FDA records medical device Class II recall Z-2761-2018, published September 26, 2018 and initiated June 21, 2018. Recalling firm: Degania Silicone, Ltd.; Status: Terminated; Reported quantity: 16133; Lot: V17018057, V17022043, V17023777, V17018058, V17012115, V17002305, V17002306, V17002308. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Degania Silicone, Ltd.
Distributor
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Lot
V17018057, V17022043, V17023777, V17018058, V17012115, V17002305, V17002306, V17002308
Risk classification
Class II
Area (FDA)
device
Country
Israel

Related recalls

Questions about this recall

Was Temperature Sensor Catheter 16FR, Catalogue Number 102201101663BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period. recalled?

Yes. FDA published recall Z-2761-2018 on September 26, 2018. Reported hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter