Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Published September 26, 2018 · Recall initiated: June 21, 2018

Summary

FDA records medical device Class II recall Z-2772-2018, published September 26, 2018 and initiated June 21, 2018. Recalling firm: Degania Silicone, Ltd.; Status: Terminated; Reported quantity: 1000; Lot: S17001781, S17002824. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Degania Silicone, Ltd.
Distributor
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Lot
S17001781, S17002824
Risk classification
Class II
Area (FDA)
device
Country
Israel

Related recalls

Questions about this recall

Was Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period. recalled?

Yes. FDA published recall Z-2772-2018 on September 26, 2018. Reported hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter