Back to search
FDAFDA · Food and drugs·Recall number Z-2756-2018

Temperature sensor Foley catheter 10FR, Catalogue Number 102205101080AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.

Published September 26, 2018 · Recall initiated: June 21, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Degania Silicone, Ltd.
Distributor
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Lot
S17005573
Risk classification
Class II
Area (FDA)
device
Country
Israel