ZOLL Pro-padz Sterile Multi-Function Electrodes (MFE) Single. Part number: 8900-4013 Product Usage: These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFEs are processed through Gamma Radiation to a SAL of 10-6 per International/ National Standards, EN ISO 11137-2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring.

Publicado el 15 de octubre de 2014 · Inicio del retiro: 16 de septiembre de 2014

Resumen

FDA registra el retiro Z-0053-2015 de dispositivos médicos de Clase II, publicado el 15 de octubre de 2014 e iniciado el 16 de septiembre de 2014. Empresa responsable: Bio-Detek, Inc.; Estado: Finalizado; Cantidad reportada: 18 singles; Lote: Lot Numbers: 1614, 1814. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Product sterility is not assured

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Bio-Detek, Inc.
Distribuidor
Worldwide Distribution - US Nationwide and the countries of: Australia, Canada, Belgium, France, and Germany.
Lote
Lot Numbers: 1614, 1814
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿ZOLL Pro-padz Sterile Multi-Function Electrodes (MFE) Single. Part number: 8900-4013 Product Usage: These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFEs are processed through Gamma Radiation to a SAL of 10-6 per International/ National Standards, EN ISO 11137-2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0053-2015 el 15 de octubre de 2014. Peligro reportado: Product sterility is not assured

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter