Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0058-2019

VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266

Publicado el 10 de octubre de 2018 · Inicio del retiro: 12 de julio de 2018

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Ortho Clinical Diagnostics Inc
Distribuidor
Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
Lote
2470 2480 2490 2501 2510 2521
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.