Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0077-2015

VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), System Product Code 6802445. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.

Publicado el 22 de octubre de 2014 · Inicio del retiro: 8 de septiembre de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Lote
Software Version 3.0 and below; Serial Numbers J46000012 - J46000456 (serial numbers are sequential; J Numbers are analogous to serial numbers)
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta